Bowers-group Sylvac D100S Readouts Uživatelský manuál Strana 58

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 74
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 57
1.10 REPLACING THE LITHIUM MODULE
If the following message appears on the screen when the unit is switched on:
This means that the data back-up lithium module is flat (average life 10 years).
It is possible to use unit D100S with a "flat" lithium module, but all parameters
and functions entered by the operator will be lost each time the unit is switched
off.
The module is replaced as follows.
- After obtaining a new lithium module from your Sylvac representative place the unit on a table and remove the 4
retaining screws of the cover.
- Touch a water pipe or other object connected to ground to release any static build-up (the inside of the unit is sen-
sitive to electrostatic discharges).
- Remove yellow cover.
- Remove lithium module with a screwdriver:
- Insert new lithium module, replace cover and replace the 4 cover retaining screws.
- When the unit is switched on again, the memory contents lost message will appear once more.
ERR 3
OUT OF MEMORY
enter one key
eine Taste drücken
presser une touche
57
Zobrazit stránku 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 73 74

Komentáře k této Příručce

Žádné komentáře